FDA Adverse Event Malfunction Summary report: N

XIENCE PRIME SV EVEROLIMUS ELUTING CORONARY STENT SYSTEM

MDR report key: 3250024 · Received July 26, 2013

Report

Report Number
2024168-2013-04684
Event Type
Malfunction
Date Received
July 26, 2013
Date of Event
June 27, 2013
Report Date
July 3, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P110019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. FAILURE TO ADVANCE/CROSS THE LESION COULD NOT BE REPLICATED IN A TESTING ENVIRONMENT AS IT WAS BASED ON OPERATIONAL CIRCUMSTANCES. SHAFT SEPARATION WAS CONFIRMED. BASED ON VISUAL AND DIMENSIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A MANUFACTURING ISSUE. IT SHOULD BE NOTED THAT THE XIENCE PRIME EVEROLIMUS ELUTING CORONARY STENT SYSTEM INSTRUCTIONS FOR USE STATES: SHOULD ANY RESISTANCE BE FELT AT ANY TIME DURING EITHER LESION ACCESS OR REMOVAL OF THE DELIVERY SYSTEM POST-STENT IMPLANTATION, THE ENTIRE SYSTEM SHOULD BE REMOVED AS A SINGLE UNIT. FAILURE TO FOLLOW THESE STEPS AND/OR APPLYING EXCESSIVE FORCE TO THE DELIVERY SYSTEM CAN POTENTIALLY RESULT IN LOSS OR DAMAGE TO THE STENT AND/OR DELIVERY SYSTEM COMPONENTS. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY;

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A DISTAL RIGHT CORONARY ARTERY. WHEN ADVANCING A 2.25 X 23 MM XIENCE PRIME RESISTANCE WAS MET AND AFTER PUSHING, THE SHAFT SEPARATED AT THE HUB. THE SEPARATED PORTION WAS REMOVED WITHOUT ISSUE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING ANOTHER XIENCE PRIME. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.