FDA Adverse Event Injury Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3249568 · Received July 26, 2013

Report

Report Number
1416980-2013-19840
Event Type
Injury
Date Received
July 26, 2013
Date of Event
July 3, 2013
Report Date
July 3, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. SHOULD RELEVANT ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

THIS IS A REPORT OF A HOME PATIENT (HP) WHO HAD A SYSTEM ERROR 2240 (AIR IN TUBING) ON THE HOMECHOICE (HC) WHILE CONNECTED DURING PERITONEAL DIALYSIS (PD) THERAPY. THERE WAS NOTHING UNUSUAL NOTED WITH THE SUPPLIES. THE PATIENT WAS PROPERLY PRIMED BEFORE CONNECTING, ALL BAGS WERE PROPERLY CONNECTED AND THERE WERE NO OPEN CLAMPS ON ANY UNUSED SUPPLY LINES. PD THERAPY WAS DISCONTINUED AND THE PATIENT DISCONNECTED FROM THE SUPPLIES. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
350027 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 63 YR HOMECHOICE