RAPID RESPONSE HUT TABLE
Report
- Report Number
- 1932056-2013-00002
- Date Received
- July 15, 2013
- Date of Event
- June 18, 2013
- Report Date
- June 19, 2013
- Manufacturer
- MEDICAL POSITIONING, INC.
- Product Code
- LGX
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THIS PRODUCT PASSED ALL OF MEDICAL POSITIONING'S PRODUCT RELEASE SAFETY INSPECTIONS AND OPERATED TO SPEC BEFORE BEING PACKAGED AND SHIPPED TO THE CUSTOMER. UPON INSPECTION AND INVESTIGATION, AFTER THE INCIDENT TOOK PLACE, IT WAS FOUND THAT THE END USER DID NOT FOLLOW ESSENTIAL LABELING PRESENT BOTH ON THE DEVICE AND IN THE PRODUCTS MANUAL. SPECIFICALLY, THE END USER DID NOT FOLLOW AN IMPORTANT ELECTRICAL WARNING WHICH STATES, "PRIOR TO CONNECTING THE PRODUCT TO ELECTRICAL POWER, CHECK TO MAKE SURE THE POWER SWITCH LOCATED ON THE LOWER POWER BOX IS IN THE "OFF" POSITION." INSPECTION OF THE PRODUCT SHOWED THAT THE WARNING LABEL WAS INTACT AND PROPERLY LOCATED ON THE PRODUCT (ATTACHED TO THE POWER CORD) AND IN THE USER INSTRUCTIONS. WHILE WE CONTINUE TO INVESTIGATE THE AFFECTED COMPONENTS, THE PRELIMINARY RESULTS INDICATE THAT THE ACT OF REPEATEDLY PLUGGING THE PRODUCT INTO AN ELECTRICAL RECEPTACLE WITH THE POWER SWITCH IN THE "ON" POSITION MAY HAVE DAMAGED THE CONTROL ELECTRONICS.
THE PT WAS UNDERGOING A SYNCOPE TEST. THE PT WAS PLACED INTO 70 DEGREE REVERSE TRENDELENBURG (HEAD-UP POSITION) ON A MEDICAL POSITIONING, INC RAPID RESPONSE HUT TABLE. AS SYMPTOMS OF SYNCOPE PRESENTED, THE NURSE ATTEMPTED TO MOVE THE PT TO A LEVEL POSITION BY PRESSING THE EMERGENCY DOWN BUTTON ON THE PRODUCT'S HAND SWITCH. THE TABLE DID NOT MOVE SO THE NURSE PRESSED THE HEAD DOWN BUTTON WHICH DIDN'T WORK EITHER. AFTER 1 MINUTE, THE PT WAS MANUALLY REMOVED FROM THE PRODUCT AND PLACED ON THE FLOOR. THE PT HAD LOST VITALS AND WAS REVIVED. ONCE STABILIZED, THE PT WAS ADMITTED INTO THE HOSPITAL, SO SHE COULD RECEIVE A PACEMAKER. DURING OUR INVESTIGATION, THE NURSE STATED THAT THE TABLE HAD BEEN "TEMPERAMENTAL" FOR A WHILE, HOWEVER, THE CUSTOMER DID NOT REPORT A COMPLAINT ON THE PRODUCT UNTIL NOW PER CUSTOMER AND MEDICAL POSITIONING, INC RECORDS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 327766 | RAPID RESPONSE HUT TABLE | RAPID RESPONSE HUT TABLE | LGX | MEDICAL POSITIONING, INC. | 1217 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | Hospitalization| L| R |