FDA Adverse Event Injury Summary report: N

EZ STAND 400 LB TO 500 LB

MDR report key: 3245045 · Received July 18, 2013

Report

Report Number
MW5030981
Event Type
Injury
Date Received
July 18, 2013
Date of Event
July 16, 2013
Report Date
July 18, 2013
Manufacturer
EZ WAY
Product Code
FNG
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
MN, US
Reporter Occupation
PATIENT

Narratives

Description of Event or Problem · 1

NURSING HOME RESIDENT WAS STANDING FROM SITTING IN A WHEELCHAIR WITH THE ASSISTANCE OF THE EZ STAND MANUFACTURED BY EZ WAY, INC. (B)(4). THERE WERE TWO CERTIFIED NURSING ASSISTANCE HELPING WITH THE EZ STAND AND RESIDENT. AS THE STAND PICKED UP THE RESIDENT OUT OF THE WHEELCHAIR, THE ACTUATOR (HYDRAULIC ARM) BROKE AT THE BOLTED POINT OF CONNECTION TO THE LIFTING ARM ON THE EZ STAND. THE LIFTING ARM OF THE EZ STAND IMMEDIATELY WENT DOWN AND DROPPED THE RESIDENT WITH A THUD INTO HER WHEELCHAIR. THE ACTUATOR FELL TOWARD THE RESIDENT WITH THE SHARP BROKEN END FACING THE RESIDENT. WE OWN THREE EZ STANDS AND UPON CHECKING THE OTHER TWO WE FOUND EACH OF THEM WERE ABOUT TO BREAK AT THE CONNECTING POINT. THIS CONNECTION IS NOT ADEQUATE FOR THE TYPE OF MOTION IT PROVIDES IN THE MECHANICAL LIFT AND WILL WEAR OUT OVER TIME WITHOUT NOTICE TO THE OPERATOR. DATES OF USE: (B)(6) 2004 - (B)(6) 2013. REASON FOR USE: ASSIST IN STANDING FROM SITTING POSITION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
335230 EZ STAND 400 LB TO 500 LB EZ STAND FNG EZ WAY

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention