NAIL HOLDING SCREW GAMMA3 8X35MM
Report
- Report Number
- 0009610622-2013-00411
- Event Type
- Malfunction
- Date Received
- July 24, 2013
- Date of Event
- July 3, 2013
- Report Date
- July 3, 2013
- Manufacturer
- STRYKER OSTEOSYNTHESIS-KIEL
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
ONCE THE INVESTIGATION HAS BEEN COMPLETED, ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.
EVALUATION CONCLUSION: EVALUATION REVEALED THAT THE NAIL HOLDING SCREW WITH LOT CODE K391539 DID NOT CONTRIBUTE TO THE COMPLAINED EVENT, THEREFORE IT WAS CLASSIFIED AS CONCOMITENT ITEM.
WE IMPLANTED THE ROD, PUT THE LAG SCREW AND DISTAL SCREW IN. ONCE EVERYTHING WAS IMPLANTED AND ATTEMPTED TO DISENGAGE, THE OUTRIGGER TO THE IMPLANT, IT WOULD NOT RELEASE. THE DOCTOR TRIED WITH ALL OF HIS STRENGTH BUT COULD NOT GET IT RELEASED. THE DISTAL LOCKING SCREW AND LAG SCREW HAD TO BE REMOVED AS WELL AS THE ROD. EVEN AFTER REMOVING THE CONSTRUCT FROM THE PATIENT, THE ROD STILL COULD NOT BE SEPARATED FROM THE INSTRUMENT. THEY BEGAN MALLET AND HAD TO FORCIBLY REMOVE THE ROD FROM THE INSTRUMENT. THE SECOND IMPLANTATION ATTEMPT WAS SUCCESSFUL.
WE IMPLANTED THE ROD, PUT THE LAG SCREW AND DISTAL SCREW IN. ONCE EVERYTHING WAS IMPLANTED AND ATTEMPTED TO DISENGAGE THE OUTRIGGER TO THE IMPLANT, IT WOULD NOT RELEASE. THE DOCTOR TRIED WITH ALL OF HIS STRENGTH BUT COULD NOT GET IT RELEASED. THE DISTAL LOCKING SCREW AND LAG SCREW HAD TO BE REMOVED AS WELL AS THE ROD. EVEN AFTER REMOVING THE CONSTRUCT FROM THE PATIENT, THE ROD STILL COULD NOT BE SEPARATED FROM THE INSTRUMENT. THEY BEGAN MALLET AND HAD TO FORCIBLY REMOVE THE ROD FROM THE INSTRUMENT. THE SECOND IMPLANTATION ATTEMPT WAS SUCCESSFUL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 345387 | NAIL HOLDING SCREW GAMMA3 8X35MM | INSTRUMENT | LXH | STRYKER OSTEOSYNTHESIS-KIEL | K391539 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 86 YR | Other |