FDA Adverse Event Malfunction Summary report: N

NAIL HOLDING SCREW GAMMA3 8X35MM

MDR report key: 3244141 · Received July 24, 2013

Report

Report Number
0009610622-2013-00411
Event Type
Malfunction
Date Received
July 24, 2013
Date of Event
July 3, 2013
Report Date
July 3, 2013
Manufacturer
STRYKER OSTEOSYNTHESIS-KIEL
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ONCE THE INVESTIGATION HAS BEEN COMPLETED, ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

EVALUATION CONCLUSION: EVALUATION REVEALED THAT THE NAIL HOLDING SCREW WITH LOT CODE K391539 DID NOT CONTRIBUTE TO THE COMPLAINED EVENT, THEREFORE IT WAS CLASSIFIED AS CONCOMITENT ITEM.

Description of Event or Problem · 1

WE IMPLANTED THE ROD, PUT THE LAG SCREW AND DISTAL SCREW IN. ONCE EVERYTHING WAS IMPLANTED AND ATTEMPTED TO DISENGAGE, THE OUTRIGGER TO THE IMPLANT, IT WOULD NOT RELEASE. THE DOCTOR TRIED WITH ALL OF HIS STRENGTH BUT COULD NOT GET IT RELEASED. THE DISTAL LOCKING SCREW AND LAG SCREW HAD TO BE REMOVED AS WELL AS THE ROD. EVEN AFTER REMOVING THE CONSTRUCT FROM THE PATIENT, THE ROD STILL COULD NOT BE SEPARATED FROM THE INSTRUMENT. THEY BEGAN MALLET AND HAD TO FORCIBLY REMOVE THE ROD FROM THE INSTRUMENT. THE SECOND IMPLANTATION ATTEMPT WAS SUCCESSFUL.

Description of Event or Problem · 1

WE IMPLANTED THE ROD, PUT THE LAG SCREW AND DISTAL SCREW IN. ONCE EVERYTHING WAS IMPLANTED AND ATTEMPTED TO DISENGAGE THE OUTRIGGER TO THE IMPLANT, IT WOULD NOT RELEASE. THE DOCTOR TRIED WITH ALL OF HIS STRENGTH BUT COULD NOT GET IT RELEASED. THE DISTAL LOCKING SCREW AND LAG SCREW HAD TO BE REMOVED AS WELL AS THE ROD. EVEN AFTER REMOVING THE CONSTRUCT FROM THE PATIENT, THE ROD STILL COULD NOT BE SEPARATED FROM THE INSTRUMENT. THEY BEGAN MALLET AND HAD TO FORCIBLY REMOVE THE ROD FROM THE INSTRUMENT. THE SECOND IMPLANTATION ATTEMPT WAS SUCCESSFUL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
345387 NAIL HOLDING SCREW GAMMA3 8X35MM INSTRUMENT LXH STRYKER OSTEOSYNTHESIS-KIEL K391539

Patients

Seq Age Sex Outcome Treatment
1 86 YR Other