FDA Adverse Event Malfunction Summary report: N

MARYLAND BIPOLAR FORCEPS INSTRUMENT

MDR report key: 3243977 · Received July 23, 2013

Report

Report Number
2955842-2013-02705
Event Type
Malfunction
Date Received
July 23, 2013
Date of Event
April 9, 2013
Report Date
June 26, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING CONFIRMED THE REPORTED COMPLAINT. THE INSTRUMENT WAS FOUND WITH A FRAYED PITCH CABLE AT THE DISTAL CLEVIS HUB. NO DAMAGE WAS FOUND AT THE CLEVIS. THE FRAYED SEGMENT WAS APPROXIMATELY 0.03 IN LENGTH. NO OTHER CABLE DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI PROSTATECTOMY PROCEDURE A SURGEON WAS COMPLAINING THAT THE GRASPER WAS NOT HOLDING TISSUE. THE SURGICAL ASSISTANT REMOVED THE FORCEPS AND A SCRUB NURSE INSPECTED IT NOTING SMALL WIRES WERE FRAYING. THERE WERE NO MISSING PIECES OR FALLEN PIECES REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME, OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
344047 MARYLAND BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420172-07 M10121022 926

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS & ACCESSORIES, ESU