FDA Adverse Event Malfunction Summary report: N

MARYLAND BIPOLAR FORCEPS INSTRUMENT

MDR report key: 3243948 · Received July 23, 2013

Report

Report Number
2955842-2013-02700
Event Type
Malfunction
Date Received
July 23, 2013
Date of Event
June 12, 2013
Report Date
June 26, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING EVALUATION FOUND THAT THE INSTRUMENT GRIP TIPS WERE BENT. THERE WAS A .033 OFFSET AT THE TIPS. ENGINEERING ALSO FOUND DEEP SCRATCHES ON THE MAIN TUBE. THE DISTAL END OF THE MAIN TUBE HAD SCRATCH MARKS EXHIBITING LIGHT MATERIAL REMOVAL AND A ROUGH SURFACE FINISH. THE SCRATCHES WERE SHORT IN LENGTH AND NOT AXIALLY ALIGNED WITH THE TUBE. THE EVIDENCE WAS INCONCLUSIVE, BUT THE DAMAGE MAY BE DUE TO MISHANDLING. NO OTHER DAMAGE WAS FOUND. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS - HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI TOTAL BENIGN HYSTERECTOMY PROCEDURE THE MARYLAND BIPOLAR FORCEPS INSTRUMENT JAWS ALLEGEDLY WOULD NOT OPEN AND CLOSE CORRECTLY. NO MISSING OR FALLEN PIECES WERE REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
342160 MARYLAND BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420172-07 M10130313 830

Patients

Seq Age Sex Outcome Treatment
1 DAVINCI SI SYSTEM, ESU, INSTRUMENTS & ACCESSORIES