FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3243926 · Received July 23, 2013

Report

Report Number
2531779-2013-11778
Event Type
Malfunction
Date Received
July 23, 2013
Report Date
July 17, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/28/2013 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP¿S BACK BOX HISTORY REVEALED TWO REBOOT EVENTS. THERE WAS A CRACK IN THE BATTERY COMPARTMENT, BUT NO EVIDENCE OF MOISTURE DAMAGE. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO POWER LOSS OR BATTERY RELATED WARNINGS PUMP CASE WAS REMOVED, NO EVIDENCE OF MOISTURE CONTAMINATION FOUND INSIDE PUMP. NO EVIDENCE OF INTERMITTENT CONTACT WAS FOUND WITH THE BATTERY CONTACTS.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS AT THE TIME OF THIS REPORT. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THEIR DEVICE WAS INTERMITTENTLY POWERING OFF AND RESTARTING. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
344491 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 16 YR