FDA Adverse Event
Malfunction
Summary report: N
RESERVOIR 3ML
MDR report key: 3243865
·
Received July 23, 2013
Report
- Report Number
- 3004209178-2013-95993
- Event Type
- Malfunction
- Date Received
- July 23, 2013
- Date of Event
- July 1, 2013
- Report Date
- July 1, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- FRN
- PMA / PMN Number
- K032005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.
Additional Manufacturer Narrative · 1
INSPECTED ONE OPEN AND USED RESERVOIR PERFORMED MANUAL PRIME AND HIGH-PRESSURE TEST PER SPECIFICATIONS. USING A NEW INFUSION SET ALONG WITH THE INSULIN PUMP, THE RESERVOIR PASSED PER SPECIFICATIONS. NO LEAKAGE ANOMALY NOTED DURING ANALYSIS. INSPECTED O-RINGS ON THE STOPPER GROOVE FOR DEFECTS AND NO ANOMALIES WERE FOUND.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT INSULIN FROM THE RESERVOIR LEAKED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 343413 | RESERVOIR 3ML | RESERVOIR | FRN | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-332A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 34 YR |