FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3243724 · Received July 23, 2013

Report

Report Number
1416980-2013-19516
Event Type
Malfunction
Date Received
July 23, 2013
Date of Event
June 28, 2013
Report Date
June 28, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AS THE SAMPLE WAS NOT AVAILABLE, AN EVALUATION COULD NOT BE PERFORMED. HOWEVER, A KNOWN CAUSE OF THE SYSTEM ERROR 2240 IS WHEN THE SUPPLY BAG DISCONNECTS WITHOUT FALLING. THE CAUSE OF THE DISCONNECTION COULD NOT BE DETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A HOMECHOICE DEVICE HAD A SYSTEM ERROR 2240 (AIR IN SET) DURING DRAIN 3 OF 5. IT WAS FURTHER REPORTED THAT THE SUPPLY BAG DISCONNECTED WITHOUT FALLING. THE PATIENT WAS CONNECTED AT THE TIME OF THE ALARM. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE MESSAGE TO THE HOME PATIENT (HP) AND INFORMED THE HP TO USE MANUAL BAGS AND CALL THE REGISTERED NURSE (RN). THERE WAS PATIENT INVOLVEMENT, BUT THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
343671 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 65 YR HOMECHOICE