FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 3243676 · Received July 23, 2013

Report

Report Number
1416980-2013-19515
Event Type
Malfunction
Date Received
July 23, 2013
Date of Event
June 27, 2013
Report Date
June 27, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). UPON INITIAL EVALUATION, THE REPORTED CONDITION WAS NOT CONFIRMED. AT THE COMPLETION OF BAXTER'S INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS SERVICED AT THE CUSTOMER SITE AND THE SAMPLE WAS VISUALLY INSPECTED AND FUNCTIONALLY TESTED. THE REPORTED CONDITION OF DAMAGED DISPLAY WAS NOT CONFIRMED. THE ALARM LOG DID NOT FIND AN ALARM ASSOCIATED WITH THE REPORTED CONDITION. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. IF ANY ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A FLOGARD INFUSION PUMP HAD A DAMAGED DISPLAY. THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS REQUESTED AND IS NOT AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
343806 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1