PUNCTUA
Report
- Report Number
- 2124215-2013-11786
- Event Type
- Malfunction
- Date Received
- July 23, 2013
- Date of Event
- June 29, 2013
- Report Date
- June 29, 2013
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT PACING THRESHOLD MEASUREMENTS ISSUES AND LOSS OF CAPTURE WERE NOTED ON BOTH LEFT AND RIGHT VENTRICLE DURING ABLATION PROCEDURE. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS PROGRAMMED AT VVI 60BPM WITH RIGHT AND LEFT VENTRICULAR OUTPUTS AT 3.5V AT 0.4MS BEFORE THE START OF THE PROCEDURE. A MINUTE AFTER RF APPLICATION, FAILURE OF VENTRICULAR CAPTURE WERE NOTED. THE CRTD WAS RE PROGRAMMED AT A HIGHER VENTRICULAR OUTPUT OF 7.5V AT 0.4MS WHICH RESOLVED THE ISSUE. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) ADVISED THAT IT IS A POTENTIAL EXPECTED BEHAVIOR DUE TO INTERACTION BETWEEN ELECTRICAL CURRENT CONDUCTED BETWEEN ELECTRODES. THIS DEVICE REMAINS IN SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 343805 | PUNCTUA | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | P053 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |