FDA Adverse Event Injury Summary report: N

EASYTRAK 2

MDR report key: 3243632 · Received July 23, 2013

Report

Report Number
2124215-2013-11813
Event Type
Injury
Date Received
July 23, 2013
Date of Event
June 26, 2013
Report Date
July 25, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
OJX
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. THE LEAD WAS RETURNED SEVERED 91 MILLIMETERS (MM) FROM THE TERMINAL PIN. VISUAL INSPECTIONS OF THE ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATION.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEFT VENTRICULAR (LV) LEAD DISLODGED. AN INVASIVE PROCEDURE WAS PERFORMED. THIS LEAD WAS EXPLANTED AND SUCCESSFULLY REPLACED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
344022 EASYTRAK 2 IMPLANTABLE LEAD OJX CPI - DEL CARIBE 4543

Patients

Seq Age Sex Outcome Treatment
1 68 YR Hospitalization| L| R 4555| N140| 0293| 4543| 4470| 4195