FDA Adverse Event Malfunction Summary report: N

TAXUS? LIBERTÉ?

MDR report key: 3243630 · Received July 23, 2013

Report

Report Number
2134265-2013-04926
Event Type
Malfunction
Date Received
July 23, 2013
Date of Event
May 16, 2013
Report Date
June 25, 2013
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
P060008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CN
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE IS COMBINATION PRODUCT. (B)(4). DEVICE EVALUATED BY MANUFACTURER: A VISUAL AND MICROSCOPIC EXAMINATION IDENTIFIED A SHAFT HYPOTUBE BREAK LOCATED 44.5 CM FROM THE CATHETER STRAIN RELIEF. THE CRIMPED STENT, BALLOON AND TIP SECTIONS OF THE DEVICE WERE VISUALLY AND MICROSCOPICALLY EXAMINED AND NO ISSUES WERE NOTED WITH THEIR PROFILES. THE BALLOON WAS TIGHTLY WRAPPED AND WAS NOT SUBJECTED TO POSITIVE PRESSURE. NO KINKS OR DAMAGE WERE NOTICED ALONG THE SHAFT OF THE DEVICE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY, AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

REPORTABLE BASED ON ANALYSIS COMPLETED ON (B)(4) 2013. IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION, SHAFT KINK OCCURRED. VASCULAR ACCESS WAS OBTAINED VIA RADIAL ARTERY. THE 32MM IN LENGTH, DE NOVO, TARGET LESION WAS LOCATED IN THE SEVERELY CALCIFIED, MODERATELY TORTUOUS LEFT ANTERIOR DESCENDING ARTERY WITH A REFERENCE DIAMETER OF 2.5MM. A 32MM X 2.50MM TAXUS LIBERTÉ STENT WAS ADVANCED BUT THE PUSHING ROD OF THE DEVICE WAS KINKED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS STABLE. HOWEVER, DEVICE ANALYSIS REVEALED HYPOTUBE BREAK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
342853 TAXUS? LIBERTÉ? CORONARY DRUG-ELUTING STENT NIQ BOSTON SCIENTIFIC - GALWAY H7493894032250 14906928

Patients

Seq Age Sex Outcome Treatment
1 65 YR