ENDOPATH** ETS FLEX45
Report
- Report Number
- 3005075853-2013-03690
- Event Type
- Malfunction
- Date Received
- July 23, 2013
- Date of Event
- June 13, 2013
- Report Date
- June 14, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K020779
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). DAMAGED FIRING TRIGGER TEETH. THE ANALYSIS RESULTS FOUND THAT DEVICE (A) WAS RECEIVED WITH THE FIRING MECHANISM DAMAGED AND WITH NO CARTRIDGE RELOAD PRESENT. NO FUNCTIONAL TEST COULD BE PERFORMED DUE TO THE CONDITION OF THE DEVICE. THE DEVICE WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS AND THE FIRING TRIGGER TEETH WERE FOUND BROKEN. WHILE NO CONCLUSION COULD BE REACH ON WHAT CAUSE THE FIRING MECHANISM TO FAIL, IT IS POSSIBLE THAT THE DEVICE WAS ATTEMPTED TO FIRE ON TICKER TISSUE THEN INDICATED OR ATTEMPTED TO FIRE TROUGH A LOCKED RELOAD IN PREVIOUS FIRINGS CAUSING AN INCREASE OF THE INTERNAL FORCES RESULTING IN THE COMPONENT YIELDING. IT SHOULD BE NOTED THAT A 100% INSPECTIONS TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRE SPECIFICATIONS; IN ADDITION, A SAMPLE OF THE BATCH IS INSPECTED AT (B)(4). THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS. THE ANALYSIS RESULTS FOUND THAT DEVICE (B) WAS RECEIVED WITH THE FIRING MECHANISM DAMAGED AND WITH NO CARTRIDGE RELOAD PRESENT. NO FUNCTIONAL TEST COULD BE PERFORMED DUE TO THE CONDITION OF THE DEVICE. THE DEVICE WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS AND THE FIRING TRIGGER TEETH WERE FOUND BROKEN. WHILE NO CONCLUSION COULD BE REACH ON WHAT CAUSE THE FIRING MECHANISM TO FAIL, IT IS POSSIBLE THAT THE DEVICE WAS ATTEMPTED TO FIRE ON TICKER TISSUE THEN INDICATED OR ATTEMPTED TO FIRE TROUGH A LOCKED RELOAD IN PREVIOUS FIRINGS CAUSING AN INCREASE OF THE INTERNAL FORCES RESULTING IN THE COMPONENT YIELDING. IT SHOULD BE NOTED THAT A 100% INSPECTIONS TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRE SPECIFICATIONS; IN ADDITION, A SAMPLE OF THE BATCH IS INSPECTED AT (B)(4). THE ANALYSIS RESULTS FOUND THAT DEVICE (C) WAS RETURNED IN GOOD VISUAL CONDITION AND WITH NO RELOAD PRESENT ON THE DEVICE. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT AND FORMED THE STAPLES AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE, THE CUT LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE TISSUE WAS NOT CUT AFTER THE DEVICE WAS FIRED. ALL THE THREE DEVICES HAVE THE SAME ISSUE. CHANGED TO ANOTHER ONE TO COMPLETE THE PROCEDURE. NO ADVERSE EVENT ON THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 342813 | ENDOPATH** ETS FLEX45 | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | J4CN29 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |