FDA Adverse Event Malfunction Summary report: N

ECHELON*FLEX60 ARTICULATING

MDR report key: 3243601 · Received July 23, 2013

Report

Report Number
3005075853-2013-03691
Event Type
Malfunction
Date Received
July 23, 2013
Date of Event
May 30, 2013
Report Date
June 21, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K081146
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION WAS NOT PROVIDED BY THE AFFILIATE. CARTRIDGE PAN DISLODGED. THE FOLLOWING INFORMATION WAS REQUESTED, BUT UNAVAILABLE: WERE THE JAWS DAMAGED? WAS THE DEVICE ABLE TO CLOSE? DID THE DEVICE HAVE TROUBLE OPENING? DID THE DEVICE CUT AS INTENDED? DID THE DEVICE STAPLE AS INTENDED? THE ANALYSIS RESULTS FOUND THAT ONE DEVICE WAS RETURNED WITH NO VISUAL NON-CONFORMANCES AND WITHOUT A CARTRIDGE RELOAD PRESENT. IN ADDITION A CARTRIDGE PAN WAS FOUND LODGED INSIDE THE CHANNEL. THE PAN WAS REMOVED FROM THE CHANNEL AND THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION WITH A TEST RELOAD. THE DEVICE FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. WHILE NO CONCLUSION COULD BE REACHED AS TO HOW THE PAN BECAME DISLODGED FROM THE RELOAD, IN ADDITION IT SHOULD BE NOTED THAT A 100% INSPECTIONS TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRE SPECIFICATIONS; IN ADDITION, A SAMPLE OF THE BATCH IS INSPECTED AT (B)(4). THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A SEGMENT RESECTION LUNG (FOR BIOPT) PROCEDURE, THE ANVIL JAW DID NOT WORK PROPERLY. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
342656 ECHELON*FLEX60 ARTICULATING STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA K4CA8F

Patients

Seq Age Sex Outcome Treatment
1