FDA Adverse Event Malfunction Summary report: N

PROMUS ELEMENT ?

MDR report key: 3243595 · Received July 23, 2013

Report

Report Number
2134265-2013-04923
Event Type
Malfunction
Date Received
July 23, 2013
Date of Event
May 13, 2013
Report Date
June 25, 2013
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
SIMILAR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE IS A COMBINATION PRODUCT. (B)(4). DEVICE EVALUATED BY MFR: A VISUAL AND MICROSCOPIC EXAMINATION OF THE DEVICE NOTED STENT DAMAGE. STRUTS THROUGHOUT THE STENT WERE SLIGHTLY MISALIGNED. NO KINKS OR DAMAGE WERE NOTICED ALONG THE SHAFT OF THE DEVICE. THE BALLOON AND TIP SECTIONS OF THE DEVICE WERE VISUALLY AND MICROSCOPICALLY EXAMINED AND NO ISSUES WERE NOTED THAT COULD HAVE POTENTIALLY CONTRIBUTED TO THE COMPLAINT INCIDENT. THE BALLOON WAS TIGHTLY WRAPPED AND WAS NOT SUBJECTED TO POSITIVE PRESSURE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

REPORTABLE BASED ON ANALYSIS COMPLETED ON (B)(4) 2013.   IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION, CROSSING DIFFICULTIES OCCURRED. THE 80% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY CALCIFIED AND SEVERELY TORTUOUS LEFT ANTERIOR DESCENDING ARTERY. A 1.5X20MM NON-BSC BALLOON CATHETER WAS USED TO PREDILATE THE LESION. THE 2.50X28MM PROMUS ELEMENT STENT WAS THEN ADVANCED AND WAS UNABLE TO CROSS THE LESION. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT' STATUS WAS STABLE.   HOWEVER, DEVICE ANALYSIS REVEALED STENT DAMAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
342441 PROMUS ELEMENT ? STENT, CORONARY, DRUG-ELUTING NIQ BOSTON SCIENTIFIC - GALWAY H7493911328250 15144715

Patients

Seq Age Sex Outcome Treatment
1 50 YR