FDA Adverse Event Malfunction Summary report: N

VOYAGER NC CORONARY DILATATION CATHETER

MDR report key: 3243440 · Received July 23, 2013

Report

Report Number
2024168-2013-04620
Event Type
Malfunction
Date Received
July 23, 2013
Date of Event
June 28, 2013
Report Date
June 28, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
LOX
PMA / PMN Number
K103110
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED INFLATION ISSUE WAS CONFIRMED. BASED ON VISUAL AND FUNCTIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LEFT MAIN ARTERY. A 3.5 X 12 MM VOYAGER NC BALLOON CATHETER WAS ADVANCED TO THE LESION; HOWEVER IT WOULD NOT INFLATE. THE BALLOON CATHETER WAS REMOVED AND A NEW 3.5 X 12 MM VOYAGER NC WAS SUCCESSFULLY USED TO COMPLETE THE PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
342448 VOYAGER NC CORONARY DILATATION CATHETER CORONARY DILATATION CATHETER LOX AV-TEMECULA-CT 2050161

Patients

Seq Age Sex Outcome Treatment
1