ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-11766
- Event Type
- Malfunction
- Date Received
- July 23, 2013
- Report Date
- July 15, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1: DATE OF SUBMISSION 10/14/2013. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/26/2013 WITH THE FOLLOWING FINDINGS: TESTING CONFIRMED INTERMITTENT RESPONSES TO BUTTON PRESSES ON THE 'UP','DOWN' AND 'CONTRAST' BUTTONS. THE 'OK' BUTTON RESPONDED TO TESTING. THERE WAS NO DAMAGE ON THE KEYPAD OBSERVED. THE KEYPAD COVER WAS REMOVED TO CHECK CONDITION OF THE BUTTON CONTACTS AND CONTAMINATION WAS OBSERVED UNDER THE CONTACTS OF ALL BUTTONS. UNRELATED TO THE INITIAL COMPLAINT, THE TEXT ON THE DISPLAY SCREEN WAS DIM AND DISCOLORED. THE DIM DISPLAY WAS REMOVED AND REPLACED WITH A TEST DISPLAY AND FOUND TO BE FULLY FUNCTIONAL WITH NO VISIBLE SIGNS OF FADING OR DISCOLORATION OF TEXT. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THE REPORTER ALLEGED THAT THE DOWN ARROW BUTTON WAS INTERMITTENTLY UNRESPONSIVE AND REQUIRED MULTIPLE PRESSES TO ELICIT A RESPONSE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 343282 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR |