STARCLOSE SE VASCULAR CLOSURE SYSTEM
Report
- Report Number
- 2024168-2013-04616
- Event Type
- Injury
- Date Received
- July 23, 2013
- Date of Event
- July 1, 2013
- Report Date
- July 2, 2013
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P050007
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4).
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE RETURNED CONDITION CONFIRMED THE REPORTED FAILURE TO ENGAGE THE DEVICE WITH THE EXCHANGE SHEATH. BASED ON THE MANUFACTURING INSPECTION CRITERIA AND ANALYSIS OF THE RETURNED DEVICE, THE PROBABLE CAUSE FOR BENT CUTTER IS INCORRECT SHEATH-TO-DEVICE INSTALLATION TECHNIQUE. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.
(B)(4). THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.
IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY WAS ATTEMPTED USING A STARCLOSE SE WITH A 6FR SHEATH, AFTER A DIAGNOSTIC PROCEDURE. REPORTEDLY, THE CLIP APPLIER OF THE STARCLOSE SE DEVICE WOULD NOT CONNECT TO THE EXCHANGE SHEATH HUB. MANUAL ARTERIAL COMPRESSION WAS APPLIED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THE PHYSICIAN IS REPORTED TO BE TRAINED IN THE USE OF THE STARCLOSE SE DEVICE. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE ENTIRE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 342588 | STARCLOSE SE VASCULAR CLOSURE SYSTEM | IMPLANTABLE CLIP | MGB | AV-TEMECULA-CT | 30315K1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | SHEATH: 6F |