FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3243214 · Received July 18, 2013

Report

Report Number
8020893-2013-01629
Event Type
Malfunction
Date Received
July 18, 2013
Date of Event
January 1, 2013
Report Date
June 19, 2013
Manufacturer
COVIDIEN, FORMERL NELLCOR
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT AN 840 VENTILATOR HAD AN ERRATIC SCREEN. NO PT INVOLVEMENT REPORTED. COVIDIEN WAS NOT AUTHORIZED TO REPAIR THE UNIT. THE CUSTOMER REPORTED TO HAVE REPLACED THE GRAPHIC USER INTERFACE (GUI) CPU PCB. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) UPDATED THE SOFTWARE AND CONDUCTED FINAL TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
336439 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERL NELLCOR 840

Patients

Seq Age Sex Outcome Treatment
1