FDA Adverse Event
Malfunction
Summary report: N
LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM 9.75CM
MDR report key: 3243118
·
Received July 18, 2013
Report
- Report Number
- 2024601-2013-00611
- Event Type
- Malfunction
- Date Received
- July 18, 2013
- Date of Event
- May 14, 2013
- Report Date
- June 25, 2013
- Manufacturer
- ALLERGAN
- Product Code
- LTI
- PMA / PMN Number
- P000008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
MEDWATCH SENT TO FDA ON: (B)(4) 2013. VISUAL EXAMINATION OF THE RETURNED DEVICE DETERMINED THE ACCESS PORT TUBING CONNECTOR TO BE A TAPER I. ALLERGAN HAS RECEIVED THE PRODUCT; HOWEVER, THE ANALYSIS HAS NOT BEEN COMPLETED AT THIS TIME. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT OR THE CONNECTOR TUBING."
Description of Event or Problem · 1
HEALTHCARE PROFESSIONAL DESCRIBES A LEAKING PORT TUBING AT THE STAINLESS STEEL CONNECTOR ON THE LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM. THE PORT WAS REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 335569 | LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM 9.75CM | LTI | ALLERGAN | NA | S99LB520 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR |