BARD INLAY LUBRICIOUS COATED DOUBLE PIGTAIL URETERAL STENT
Report
- Report Number
- 1018233-2013-02912
- Event Type
- Malfunction
- Date Received
- July 18, 2013
- Date of Event
- June 4, 2013
- Report Date
- June 10, 2013
- Manufacturer
- PRODUCTOS PARA EL CUIDADO DE LA SALUD
- Product Code
- FAD
- PMA / PMN Number
- K983498
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- PHYSICIAN
Narratives
THE COMPLAINT IS CONFIRMED WITH CAUSE UNK. THE RETURNED STENT WAS BROKEN FROM ONE OF THE PIGTAILS. THERE WAS NO SUTURE OR PIGTAIL RETURNED FOR EVAL. THE PACKAGE WAS RECEIVED OPENED AND THE INTERNAL UNIT PACKAGING (PLASTIC BAG) WAS OPENED. THE BREAKAGE SITE PRESENTS JAGGED EDGES AND STRESS MARKS AS INDICATIVE THAT THE STENT HAS BEEN STRESSED BEYOND ITS TENSILE CAPABILITIES. THE BROKEN SECTION IS LOCATED AT ONE OF THE HOLES/PERFORATIONS OF PIGTAIL STENT. THE LENGTH OF STENT IS 10.500". DIMENSIONAL EVAL REVEALED THAT THE STENT WAS MANUFACTURED ACCORDING TO SPECIFICATIONS. AN ENTIRE REVIEW OF DEVICE HISTORY RECORDS WAS DONE FOR THE INVOLVED LOT AND NO DISCREPANCIES WERE FOUND THAT WOULD HAVE CONTRIBUTED TO THE REPORTED ISSUE. THERE WERE NO DOCUMENTED INTERNAL REJECTS RELATED TO THE REPORTED ISSUE DURING THE MANUFACTURING OF THE REPORTED PRODUCTION BATCH. THE INSTRUCTIONS FOR USE STATE THE FOLLOWING: IMPROPER HANDLING TECHNIQUE CAN SERIOUSLY WEAKEN THE STENT. ACUTE BENDING OR OVERSTRESSING DURING PLACEMENT COULD RESULT IN SUBSEQUENT SEPARATION OF THE STENT AT THE POINT OF STRESS AFTER A PROLONGED INDWELLING PERIOD. SUTURE MAY BE CUT OFF PRIOR TO STENT PLACEMENT. REMOVE SUTURE PRIOR TO PLACEMENT FOR PEDIATRIC PTS. EXERCISE CARE. TEARING OF THE STENT CAN BE CAUSED BY SHARP INSTRUMENTS. CARE SHOULD BE EXERCISED WHEN REMOVING STENT FROM INNER POLYBAG SO AS NOT TO CAUSE TEARING OR FRAGMENTATION. (B)(4).
(B)(4). IT WAS REPORTED THAT ONE EXTREMITY OF THIS DOUBLE PIGTAIL URETERAL STENT PRESENTED DEFECT. PER ADDITIONAL INFORMATION RECEIVED, THE PIGTAIL END OF THE STENT WAS MISSING AND NO BROKEN PIECES WERE FOUND IN THE PACKAGING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 335311 | BARD INLAY LUBRICIOUS COATED DOUBLE PIGTAIL URETERAL STENT | FAD | PRODUCTOS PARA EL CUIDADO DE LA SALUD | NA | NGWJ1768 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |