FDA Adverse Event Malfunction Summary report: N

11CM ANGLE ATTACHMENT

MDR report key: 3242803 · Received July 23, 2013

Report

Report Number
1045834-2013-03588
Event Type
Malfunction
Date Received
July 23, 2013
Report Date
November 14, 2011
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
K011444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED BY DEPUY SYNTHES POWER TOOLS. RELIABILITY ENGINEERING EVALUATED THE DEVICE, AND THE REPORTED PROBLEM WAS NOT CONFIRMED; A CUTTING BUR COULD BE INSERTED AND LOCKED INTO THE ATTACHMENT. HOWEVER, THE BEARINGS AND GEARS IN THE ELBOW AND BASE WERE CORRODED AND DAMAGED, AND THE UNIT WOULD NOT ROTATE. THIS CONDITION IS INDICATIVE OF IMPROPER OR INEFFECTIVE CLEANING AND MAINTENANCE OF THE EQUIPMENT. IF ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SENT. (B)(4).

Description of Event or Problem · 1

REPORT RECEIVED FROM (B)(6) STATING THAT THE LOCKING MECHANISM IS DAMAGED. THE DEVICE WAS NOT BEING USED DURING SURGERY. IT IS UNKNOWN IF INJURY OR MEDICAL INTERVENTION OCCURRED. THE DATE OF THE EVENT IS UNKNOWN. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
343178 11CM ANGLE ATTACHMENT MOTOR, DRILL, ELECTRIC - ATTACHMENT HBC DEPUY SYNTHES POWER TOOLS