11CM ANGLE ATTACHMENT
Report
- Report Number
- 1045834-2013-03588
- Event Type
- Malfunction
- Date Received
- July 23, 2013
- Report Date
- November 14, 2011
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HBC
- PMA / PMN Number
- K011444
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TU
- Reporter Occupation
- OTHER
Narratives
THE DEVICE WAS RECEIVED BY DEPUY SYNTHES POWER TOOLS. RELIABILITY ENGINEERING EVALUATED THE DEVICE, AND THE REPORTED PROBLEM WAS NOT CONFIRMED; A CUTTING BUR COULD BE INSERTED AND LOCKED INTO THE ATTACHMENT. HOWEVER, THE BEARINGS AND GEARS IN THE ELBOW AND BASE WERE CORRODED AND DAMAGED, AND THE UNIT WOULD NOT ROTATE. THIS CONDITION IS INDICATIVE OF IMPROPER OR INEFFECTIVE CLEANING AND MAINTENANCE OF THE EQUIPMENT. IF ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SENT. (B)(4).
REPORT RECEIVED FROM (B)(6) STATING THAT THE LOCKING MECHANISM IS DAMAGED. THE DEVICE WAS NOT BEING USED DURING SURGERY. IT IS UNKNOWN IF INJURY OR MEDICAL INTERVENTION OCCURRED. THE DATE OF THE EVENT IS UNKNOWN. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 343178 | 11CM ANGLE ATTACHMENT | MOTOR, DRILL, ELECTRIC - ATTACHMENT | HBC | DEPUY SYNTHES POWER TOOLS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |