FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3242525 · Received July 23, 2013

Report

Report Number
1416980-2013-19426
Event Type
Malfunction
Date Received
July 23, 2013
Date of Event
June 28, 2013
Report Date
June 28, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION, LEAK TESTING, CLEAR PASSAGE TESTING, AND CLAMP FUNCTION TESTING WERE PERFORMED WITH NO ISSUES NOTED. THE REPORTED ISSUE COULD NOT BE CONFIRMED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING TROUBLESHOOTING FOR AN UNRELATED ALARM ON A HOMECHOICE (HC) MACHINE, THE HOME PATIENT (HP) STATED THAT THERE WAS A STRAIGHT WHITE LINE IN THE WALL OF THE TUBING OF THE PATIENT LINE. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ADVISED THE HOME PATIENT (HP) TO START OVER WITH NEW SUPPLIES. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED IN ASSOCIATION WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
343227 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 79 YR HOMECHOICE