ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-11702
- Event Type
- Malfunction
- Date Received
- July 23, 2013
- Report Date
- June 24, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED A DIM, DISCOLORED DISPLAY SCREEN. A TEST SCREEN WAS USED FOR INVESTIGATION AND WAS FOUND TO FUNCTION PROPERLY. UNRELATED TO THE DISPLAY ISSUE, TESTING REVEALED THAT THE TIME AND DATE RESET TO FACTORY SETTINGS. THE PUMP WAS OPENED AND EVALUATION REVEALED THE INTERNAL CLOCK BATTERY ON THE PRINTED CIRCUIT BOARD WAS LEAKING. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THIS REPORT IS MADE UNDER THE REQUIREMENTS OF THE MEDICAL DEVICE REPORTING REGULATIONS AND DOES NOT CONSTITUTE AN ADMISSION ON THE PART OF ANIMAS OF ANY DEFICIENCY IN THE PERFORMANCE OF THE DEVICE. (B)(6).
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE DISPLAY SCREEN WAS DIM AND DISCOLORED. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2013.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 344270 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 22 YR |