ACCU-CHEK ® SPIRIT
Report
- Report Number
- 2183996-2013-01358
- Event Type
- Malfunction
- Date Received
- July 23, 2013
- Date of Event
- June 27, 2013
- Report Date
- January 23, 2014
- Manufacturer
- ROCHE HEALTH SOLUTIONS, INC.
- Product Code
- LZG
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
AS THE PRODUCT HAS NOT BEEN RETURNED FOR INVESTIGATION AND THE PRODUCTION DATA COMPLY WITH THE SPECIFICATION, THE COMPLAINT COULD NOT BE REPLICATED. PRODUCTION REPORTS WERE REVIEWED. DEVICE WAS NOT RETURNED TO MANUFACTURER.
THE COMPLAINT CANNOT BE VERIFIED DUE TO CARELESS HANDLING OF THE PRODUCT. THE SOFT COMPONENTS OF THE HOUSING ARE WORN DOWN HEAVILY. THEREFORE, MOISTURE MAY ENTER THE INSULIN PUMP AND DESTROY THE PUMP ELECTRONICS. THE FUNCTIONALITY OF THE BUTTONS WAS TESTED SUCCESSFULLY.
THIS INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED IT WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT.
PATIENT REPORTED THE PROTECTIVE RUBBER FELL OFF THE TOP BUTTON AND IS WORN ON THE SIDE BUTTON, AND IT OFTEN TAKES MORE THAN 1 ATTEMPT FOR THE BUTTONS TO RESPOND. THE INFUSION DEVICE WAS REPLACED AND REQUESTED FOR EVALUATION. NO PHYSIOLOGICAL EFFECTS WERE REPORTED, AND SHE DID NOT REQUIRE TREATMENT FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 343029 | ACCU-CHEK ® SPIRIT | INSULIN INFUSION PUMP | LZG | ROCHE HEALTH SOLUTIONS, INC. | 00700006862 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 030 YR |