ACCU-CHEK ® SPIRIT COMBO
Report
- Report Number
- 2183996-2013-01353
- Event Type
- Malfunction
- Date Received
- July 23, 2013
- Date of Event
- July 11, 2013
- Report Date
- August 21, 2013
- Manufacturer
- ROCHE HEALTH SOLUTIONS INC
- Product Code
- LZG
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
E7002 WERE FOUND IN THE HISTORY. THE VIBRATOR MOTOR DOES NOT FUNCTION. AN INTERNAL DEFECT OF THE VIBRATOR LED TO THE PROBLEM. THE ACOUSTIC AND VISUAL ALARM FUNCTIONS ARE NOT AFFECTED.
THE INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED, IT WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT.
ON (B)(6) 2013, IT WAS REPORTED THAT THE PATIENTS INFUSION DEVICE HAD DISPLAYED E7 (ELECTRONIC ERROR) AND E8 (POWER INTERRUPT). IT WAS REPORTED THAT THE VIBRA-ALARM WAS NOT FUNCTIONING ON OCCASION. NO PHYSIOLOGICAL EFFECTS WERE REPORTED. THE PATIENT DID NOT REQUIRE MEDICAL ASSISTANCE FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE. THE INFUSION DEVICE WAS REQUESTED TO BE RETURNED FOR PRODUCT EVALUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 342917 | ACCU-CHEK ® SPIRIT COMBO | INSULIN INFUSION PUMP | LZG | ROCHE HEALTH SOLUTIONS INC | 00700006863 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 048 YR |