FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3242429 · Received July 23, 2013

Report

Report Number
2531779-2013-11682
Event Type
Malfunction
Date Received
July 23, 2013
Report Date
July 14, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 18-JUL-2019 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, THE PUMP WAS POWERED WITH A CLEAR AND LEGIBLE DISPLAY. THE COMPLAINT OF A BLANK DISPLAY WAS NOT DUPLICATED. UNRELATED TO THE COMPLAINT, INVESTIGATION REVEALED A DIM, DISCOLORED DISPLAY. THE PUMP WAS OPENED AND NO EVIDENCE OF MOISTURE WAS OBSERVED INSIDE THE PUMP. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (BLANK SCREEN) ISSUE. THE REPORTER ALLEGED THAT AFTER THE BATTERY WAS CHANGED, THE DISPLAY SCREEN ON THE PUMP REMAINED BLANK AND THE REPORTER NOTED AUDIBLE TONES/VIBRATIONS. IT WAS REPORTED THAT THE PATIENT HAD BLOOD GLUCOSE (BG) OF 445MG/DL WITHOUT SYMPTOMS ON THE MORNING OF THE EVENT. THE REPORTED BG WAS NOT INDICATIVE OF A SERIOUS INJURY AND DOES NOT MEET THE ANIMAS CRITERIA OF AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
343013 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 71 YR