FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3242225 · Received July 23, 2013

Report

Report Number
2531779-2013-11667
Event Type
Malfunction
Date Received
July 23, 2013
Report Date
July 11, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

IF ANIMAS RECEIVES THE PRODUCT ANIMAS WILL INFORM THE FDA OF ANY PRODUCTS THAT FAIL EVALUATION IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 11/14/2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP¿S BLACK BOX HISTORY SHOWED THAT THE DATA FROM THE DATE OF THE REPORTED EVENT WAS OVERWRITTEN DUE TO CONTINUED USE OF THE PUMP. NO ACTIVITY OUTSIDE OF NORMAL USE WAS OBSERVED IN THE BLACK BOX AND DOWNLOAD HISTORY. THE RETURNED BATTERY CAP WAS ABLE TO BE SECURED TO THE PUMP AND WAS USED TO COMPLETE ALL TESTING. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO ALARMS OR POWER ISSUES OCCURRING. THE PUMP CASE WAS REMOVED AND NO DAMAGE WAS FOUND TO THE POWER CIRCUIT. THE COMPLAINT THAT THE PUMP HAD NO POWER WAS NOT ABLE TO BE DUPLICATED DURING INVESTIGATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (NO POWER) ISSUE. THE REPORTER STATED THAT SHE DID NOT CLEAR THE ALARMS ON HER PUMP AND WHEN SHE WENT TO REBOOT THE PUMP, SHE WAS UNABLE TO POWER IT ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
342979 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 52 YR