FDA Adverse Event Malfunction Summary report: N

MICRO DRILL MEDIUM STRAIGHT ATTACHMENT

MDR report key: 3242212 · Received July 23, 2013

Report

Report Number
0001811755-2013-01748
Event Type
Malfunction
Date Received
July 23, 2013
Date of Event
June 20, 2013
Report Date
June 26, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FOLLOW UP REPORT WILL BE FILED AFTER THE QUALITY INVESTIGATION HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE REPORTED EVENT OF OVERHEATING WAS CONFIRMED DURING THE DEVICE EVALUATION. UPON VISUAL INSPECTION, SEVERE CORROSION WAS FOUND TO BE BUILT UP WITHIN THE ATTACHMENT AND THE UPPER BEARING WAS FOUND TO BE SHATTERED, WHICH IS THE PROBABLE CAUSE OF THE REPORTED OVERHEATING. THE DEVICE WAS DISCARDED BY THE MANUFACTURER.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, EXTRACTION OF TOOTH NO. 8, AT THE USER FACILITY THE MICRO DRILL MEDIUM STRAIGHT ATTACHMENT WAS OVERHEATING. IT WAS REPORTED THAT THE PATIENT WAS BURNED AT THE LIP AND TREATED WITH AN OINTMENT AND PER THE DOCTOR ¿THE PATIENT BECAME WELL¿. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE SAME DEVICE WITH NO DELAY.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, EXTRACTION OF TOOTH NO. 8, AT THE USER FACILITY THE MICRO DRILL MEDIUM STRAIGHT ATTACHMENT WAS OVERHEATING. IT WAS REPORTED THAT THE PATIENT WAS BURNED AT THE LIP AND TREATED WITH AN OINTMENT AND PER THE DOCTOR ¿THE PATIENT BECAME WELL¿. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE SAME DEVICE WITH NO DELAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
342363 MICRO DRILL MEDIUM STRAIGHT ATTACHMENT DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE ERL STRYKER INSTRUMENTS-KALAMAZOO 10302

Patients

Seq Age Sex Outcome Treatment
1