FDA Adverse Event Malfunction Summary report: N

LIFEPAK CR(R) PLUS DEFIBRILLATOR

MDR report key: 3242186 · Received July 23, 2013

Report

Report Number
3015876-2013-00645
Event Type
Malfunction
Date Received
July 23, 2013
Date of Event
June 25, 2013
Report Date
June 25, 2013
Manufacturer
PHYSIO-CONTROL, INC
Product Code
MKJ
PMA / PMN Number
K033275
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL CONTINUES TO INVESTIGATE THE REPORTED FAILURE AND WILL SUBMIT A SUPPLEMENTAL REPORT ON THIS EVENT TO THE FDA AS PROVIDED BY 21 CFR 803.56.

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO DETERMINED THAT THE CAUSE OF THE REPORTED FAILURE WAS DUE TO THREE ELECTRICALLY LEAKY FILTERS, DESIGNATORS FL9, FL10 AND FL11 LOCATED ON THE ANALOG PCB ASSEMBLY. THE CUSTOMER RECEIVED A REPLACEMENT DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD THE THREE ICONS (CHARGE PAK, ATTENTION AND SERVICE WRENCH) ILLUMINATED ON THE READINESS DISPLAY. THEREFORE THE DEVICE WOULD LIKELY NOT HAVE ENOUGH POWER TO PROVIDE ADEQUATE DEFIBRILLATION THERAPY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
344242 LIFEPAK CR(R) PLUS DEFIBRILLATOR DEFIBRILLATORS, AUTOMATIC, EXTERNAL MKJ PHYSIO-CONTROL, INC CRPLUS

Patients

Seq Age Sex Outcome Treatment
1