FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK CR(R) PLUS DEFIBRILLATOR
MDR report key: 3242186
·
Received July 23, 2013
Report
- Report Number
- 3015876-2013-00645
- Event Type
- Malfunction
- Date Received
- July 23, 2013
- Date of Event
- June 25, 2013
- Report Date
- June 25, 2013
- Manufacturer
- PHYSIO-CONTROL, INC
- Product Code
- MKJ
- PMA / PMN Number
- K033275
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). PHYSIO-CONTROL CONTINUES TO INVESTIGATE THE REPORTED FAILURE AND WILL SUBMIT A SUPPLEMENTAL REPORT ON THIS EVENT TO THE FDA AS PROVIDED BY 21 CFR 803.56.
Additional Manufacturer Narrative · 1
PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO DETERMINED THAT THE CAUSE OF THE REPORTED FAILURE WAS DUE TO THREE ELECTRICALLY LEAKY FILTERS, DESIGNATORS FL9, FL10 AND FL11 LOCATED ON THE ANALOG PCB ASSEMBLY. THE CUSTOMER RECEIVED A REPLACEMENT DEVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE HAD THE THREE ICONS (CHARGE PAK, ATTENTION AND SERVICE WRENCH) ILLUMINATED ON THE READINESS DISPLAY. THEREFORE THE DEVICE WOULD LIKELY NOT HAVE ENOUGH POWER TO PROVIDE ADEQUATE DEFIBRILLATION THERAPY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.