HOMECHOICE
Report
- Report Number
- 1416980-2013-19381
- Event Type
- Malfunction
- Date Received
- July 23, 2013
- Date of Event
- January 20, 2013
- Report Date
- June 27, 2013
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS IS AN ANCILLARY SERVICE EVENT. THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS ONE OR MORE CYCLE ADVANCES TO THE NEXT FILL WHEN A SLOW / NO FLOW OCCURRED, ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW-UP REPORT WILL BE SENT.
DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED. THIS OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2013, DURING NIGHT DRAIN CYCLE. THE PATIENT'S ULTRAFILTRATION READING WAS 1315ML, INDICATING THE HOME PATIENT (HP) DRAINED 1315ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 343703 | HOMECHOICE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |