FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 3241980 · Received July 23, 2013

Report

Report Number
1416980-2013-19381
Event Type
Malfunction
Date Received
July 23, 2013
Date of Event
January 20, 2013
Report Date
June 27, 2013
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS IS AN ANCILLARY SERVICE EVENT. THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS ONE OR MORE CYCLE ADVANCES TO THE NEXT FILL WHEN A SLOW / NO FLOW OCCURRED, ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW-UP REPORT WILL BE SENT.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED. THIS OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2013, DURING NIGHT DRAIN CYCLE. THE PATIENT'S ULTRAFILTRATION READING WAS 1315ML, INDICATING THE HOME PATIENT (HP) DRAINED 1315ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
343703 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1