FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3241979 · Received July 23, 2013

Report

Report Number
1416980-2013-19380
Event Type
Malfunction
Date Received
July 23, 2013
Date of Event
May 8, 2013
Report Date
June 28, 2013
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS EVALUATED AND THE REPORTED ISSUE WAS CONFIRMED THROUGH THE REVIEW OF THE LOGS. THE CAUSE WAS DETERMINED TO BE USE ERROR, TIDAL TOTAL ULTRA FILTRATION (UF) REMOVAL WAS SET TOO LOW. THE HOMECHOICE APD SYSTEMS TRAINER'S GUIDE GIVES INSTRUCTIONS ON HOW TO SET THE TIDAL THERAPY SETTINGS. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING THE EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED, WHICH OCCURRED DURING THE THERAPY INITIATED ON (B)(6) 2013 AT 01:06:22. DURING NIGHT DRAIN CYCLE NINE, THE PATIENT'S ULTRAFILTRATION READING WAS 1760ML, INDICATING THE HOME PATIENT (HP) DRAINED 1430ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2200ML. NO ADDITIONAL INFORMATION IS AVAILABLE.