UV FLASH TRANSFER SETS
Report
- Report Number
- 1416980-2013-19384
- Event Type
- Malfunction
- Date Received
- July 23, 2013
- Date of Event
- June 27, 2013
- Report Date
- June 27, 2013
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- KDJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NURSE
Narratives
(B)(4). THE DEVICE WAS RETURNED FOR EVALUATION. A REVIEW OF ALL BATCH RECORD DOCUMENTS WAS PERFORMED WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE AND NO EXCEPTIONS RELATED TO THE REPORTED CONDITION WERE NOTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
(B)(4). THREE USED DEVICES WERE RECEIVED. VISUAL INSPECTION FOUND TWO SETS TO HAVE FLARED PATIENT ADAPTERS. LEAK TESTING, CLEAR PASSAGE TESTING, AND CLAMP FUNCTION TESTING WERE PERFORMED WITH NO ISSUES NOTED. THE INTEGRITY OF THE SEAL SURFACE WAS TESTED WITH NO LEAKS NOTED. THE INSIDE DIAMETER OF THE PATIENT CONNECTOR WAS MEASURED AND FOUND TO BE BELOW NOMINAL. IMPROVEMENTS WERE MADE TO THE MOLD AND THE MOLDING DEPARTMENT PROCEDURES TO ADDRESS THIS ISSUE. (B)(4). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
(B)(4). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT THE PATIENT CONNECTOR OF A UV FLASH TRANSFER SET COULD NOT BE FIRMLY CONNECTED WITH THE TITANIUM ADAPTER AT THE TIME OF PLACEMENT. THERE WAS NO ADVERSE EVENT REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 342366 | UV FLASH TRANSFER SETS | SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE | KDJ | BAXTER HEALTHCARE - MOUNTAIN HOME | H13A02017 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | TITANIUM ADAPTOR |