FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3241949 · Received July 22, 2013

Report

Report Number
2531779-2013-11662
Event Type
Malfunction
Date Received
July 22, 2013
Report Date
July 19, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 10/10/2013 DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/20/2013 WITH THE FOLLOWING FINDINGS: THE PUMP WAS RETURNED WITH TORN KEYPAD ABOVE THE ¿UP¿ KEY. ALL KEYPAD BUTTONS WERE COMPLETELY UNRESPONSIVE. WHEN THE KEYPAD WAS REMOVED TO INVESTIGATE, EVIDENCE OF KEYPAD CONTAMINATION WAS LOCATED, UNDER ALL KEYPAD BUTTONS.

Additional Manufacturer Narrative · 1

THE PRODUCTS HAVE NOT YET BEEN RETURNED. IF THE PRODUCT(S) ARE RETURNED, ANM WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING AN UNRESPONSIVE KEYPAD ISSUE. ONE WEEK PRIOR TO THE COMPLAINT, THE "OK" AND "DOWN" BUTTONS BECAME INTERMITTENTLY RESPONSIVE. THE DAY BEFORE THE COMPLAINT, THE ISSUE PROGRESSED AND THE TWO BUTTONS BECAME COMPLETELY UNRESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
341026 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 44 YR