ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-11655
- Event Type
- Malfunction
- Date Received
- July 22, 2013
- Report Date
- July 18, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP #1: DATE OF SUBMISSION 10/08/2013 DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/19/2013 WITH THE FOLLOWING FINDINGS: THE KEYPAD IS TORN FROM THE UP TO THE OK KEYPAD SYMBOLS. THE CONTRAST BUTTON HAS AN INTERMITTENT RESPONSE. THE UP, DOWN, & OK BUTTONS RESPOND PROPERLY. CONTAMINATION WAS FOUND UNDER THE CONTRAST AND DOWN BUTTONS WHEN THE KEYPAD WAS REMOVED. UNRELATED TO THE COMPLAINT, INVESTIGATION REVEALED THAT WHEN BOOTED TO THE VERIFY SCREEN, THE PUMP HAD A DIM PINK CONTRAST. THE PUMP COVER WAS REMOVED AND THE OLED SCREEN WAS REMOVED AND REPLACED WITH A NEW TEST SCREEN. ONCE COMPLETED, THE CONTRAST RETURNED TO NORMAL.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGE PRIOR TO DAMAGE) ISSUE. IT WAS REPORTED THAT THERE WAS A TEAR FROM TOP OF UP ARROW TO THE OK BUTTON, ALONG WITH UNRESPONSIVENESS OF BUTTONS FOR A COUPLE HOURS SEVERAL WEEKS AGO. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 340183 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 13 YR |