FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3241881 · Received July 22, 2013

Report

Report Number
2531779-2013-11648
Event Type
Malfunction
Date Received
July 22, 2013
Report Date
July 18, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/29/2013 WITH THE FOLLOWING FINDINGS: MOISTURE WAS OBSERVED BEHIND THE DISPLAY LENS. THE PUMP POWERED ON SUCCESSFULLY TO THE VERIFY SCREEN. MULTIPLE POWER ON RESETS AND REBOOTS WERE OBSERVED IN A REVIEW OF THE DEVICE HISTORY. THE AUDIO BOLUS BUTTON COVER WAS MISSING, AND A LEAK TEST INDICATED A BOLUS BUTTON LEAK. THE PUMP CASE WAS OPENED AND EVIDENCE OF MOISTURE DAMAGE WAS OBSERVED ON MULTIPLE INTERNAL COMPONENTS.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (MOISTURE INGRESS) ISSUE. IT WAS REPORTED THAT THE PUMP HAS MOISTURE IN THE DISPLAY SCREEN. THE INFORMATION PROVIDED TO ANIMAS INDICATED THAT THE PATIENT WENT SWIMMING AND HAD THE ISSUE AFTERWARDS. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
341495 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 14 YR