FDA Adverse Event Malfunction Summary report: N

COULTER® HMX HEMATOLOGY ANALYZER WITH AUTOLOADER

MDR report key: 3241796 · Received July 22, 2013

Report

Report Number
1061932-2013-01532
Event Type
Malfunction
Date Received
July 22, 2013
Date of Event
June 25, 2013
Report Date
June 25, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K010765
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) OBSERVED THE PROBE SECTION OF THE BLOOD SAMPLING VALVE (BSV) WAS LEAKING AND PINCH VALVE PV42 TUBING WAS NOT PROPERLY INSERTED INTO THE PINCH VALVE. THE FSE RESEATED THE TUBING AND RESOLVED THE FLUID LEAK ISSUE. SERVICE ACTIVITY PERFORMED WAS VERIFIED TO MEET THE SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURES. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED PERFORMANCE SPECIFICATIONS AND WAS RETURNED TO NORMAL OPERATION. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED APPROXIMATELY ONE MILLILITER OF CLEAR BLOOD FLUID LEAKED BELOW THE ASPIRATION NEEDLE AND CONDUCTIVITY WAS ELEVATED AND OUT OF SPECIFICATION DURING QUALITY CONTROL (QC) ANALYSIS, INVOLVING THE COULTER HMX HEMATOLOGY ANALYZER WITH AUTOLOADER. THE FLUID WAS CONTAINED WITHIN THE INSTRUMENT. THE OPERATOR WAS WEARING PROPER PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES, A LABORATORY COAT, AND EYE PROTECTION AND DID NOT HAVE DIRECT CONTACT WITH THE FLUID. THERE WAS NO OPERATOR INJURY OR ADVERSE EFFECT ASSOCIATED WITH THIS EVENT. PATIENT RESULTS WERE NOT GENERATED AS THE EVENT OCCURRED DURING QUALITY CONTROL ANALYSIS. THERE WAS NO PATIENT CONSEQUENCE. THE FACILITY HAS AN EXPOSURE CONTROL AND RISK MANAGEMENT PLAN IN PLACE FOR BIOHAZARD MATERIAL. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ASSESS THE INSTRUMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340558 COULTER® HMX HEMATOLOGY ANALYZER WITH AUTOLOADER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1