OLYMPUS AU481-02E CLINICAL CHEMISTRY ANALYZER (WITH ISE)
Report
- Report Number
- 9612296-2013-00096
- Event Type
- Malfunction
- Date Received
- July 22, 2013
- Date of Event
- June 24, 2013
- Report Date
- June 28, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- JGS
- PMA / PMN Number
- K981743
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
SAMPLE COLLECTION AND CENTRIFUGATION INFORMATION WAS NOT PROVIDED. QUALITY CONTROL (QC) WAS RUN PRIOR TO AND AFTER THE EVENT AND WAS OUT OF THE LABORATORY SPECIFIED RANGES. THE CUSTOMER DID NOT PROVIDE ANY DOCUMENTATION OF CALIBRATION OR QC DATA. SERVICE WAS REQUESTED. A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED FOR THIS EVENT. THE FSE EVALUATED THE IONIZED SELECTIVE ELECTRODE (ISE) SYSTEM AND RESOLVED THE ISSUE BY REPLACING THE ELECTRODE WIRE CABLES / LEADS. THE FSE REVIEWED THE CALIBRATION SLOPE VALUES AND NOTED BETTER RECOVERY AND STABILITY. THE CUSTOMER VERIFIED ISE CALIBRATION RESULTS WERE IMPROVED. SYSTEM PERFORMANCE WAS VERIFIED TO MEET PUBLISHED SPECIFICATIONS. THE CUSTOMER ALSO VERIFIED ALL QC LEVELS WERE WITHIN INTERNAL FACILITY SPECIFICATIONS BEFORE RUNNING PATIENT SAMPLES. THE CUSTOMER CONFIRMED THAT THEY HAVE HAD NO FURTHER ISE ISSUES WITH CALIBRATION DRIFT OR LOW ISE VALUES. THE FAILURE MODE IS RELATED TO HARDWARE, ISE ELECTRODE LEADS. MDR 9612296-2013-00095 IS RELATED TO THIS EVENT WHICH DOCUMENTS THE ERRONEOUSLY LOW NA PATIENT RESULTS OBTAINED ON (B)(6) 2013.
A CUSTOMER CONTACTED BECKMAN COULTER (BEC) IN REGARDS TO OBTAINING ERRONEOUSLY LOW SODIUM (NA) PATIENT RESULTS GENERATED ON THE (B)(4) CLINICAL CHEMISTRY ANALYZER ON TWO DIFFERENT DAYS (B)(6) 2013. THIS REPORT DOCUMENTS THE THREE (3) ERRONEOUSLY LOW NA PATIENT RESULTS OBTAINED ON (B)(6) 2013. THE CUSTOMER ALSO REPORTED ONGOING CALIBRATION AND QUALITY CONTROL (QC) INSTABILITY AND DRIFT ISSUES ON THE ANALYZER. NO ERRONEOUS RESULTS WERE REPORTED OUT OF THE LABORATORY. THE CUSTOMER DID NOT PROVIDE SPECIFIC PATIENT DEMOGRAPHICS OR DATA DOCUMENTATION. THERE WAS NO AFFECT TO PATIENT OR PATIENT TREATMENT IN CONNECTION TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 340457 | OLYMPUS AU481-02E CLINICAL CHEMISTRY ANALYZER (WITH ISE) | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JGS | BECKMAN COULTER | AU481-02E (WITH ISE) | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |