COULTER® LH 750 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2013-01568
- Event Type
- Malfunction
- Date Received
- July 22, 2013
- Date of Event
- June 27, 2013
- Report Date
- June 27, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K061574
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE CUSTOMER RE-ATTACHED THE TUBING THAT HAD POPPED OFF THE BLOOD SAMPLING VALVE (BSV). THE CUSTOMER INDICATED THAT THE LEAK WAS STILL PRESENT. A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT. THE FSE REROUTED THE TUBING ON THE BSV TO GIVE IT MORE PLAY DURING THE ROTATION OF THE BSV. THE FSE ALSO TIGHTENED THE PROBE WIPE WASTE TRUCK WHICH WAS LOOSE. THE FSE VERIFIED THE INSTRUMENT WITH NO FURTHER LEAKS OBSERVED. PER LABELING, BECKMAN COULTER URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IT IS NOT LIMITED TO, PROTECTIVE EYEWEAR, GLOVES, AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER. (B)(4).
A CUSTOMER CONTACTED BECKMAN COULTER (BEC) REPORTING 50 CUBIC CENTIMETERS (CCS) OF DILUENT THAT LEAKED ONTO THE COUNTER TOP FROM THE COULTER LH 750 HEMATOLOGY ANALYZER. THE CUSTOMER STATED THAT THEY HAD ALREADY PERFORMED STARTUP, CONTROLS, AND RUN ABOUT FIFTEEN (15) PATIENT SAMPLES FOR THE DAY WHEN THE LEAK WAS NOTICED. THE CUSTOMER OPENED THE FRONT OF THE INSTRUMENT AND FOUND THAT THE LEAK WAS COMING FROM THE TUBING THAT HAD POPPED OFF OF THE BLOOD SAMPLING VALVE (BSV). THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES AND A LABORATORY COAT AT THE TIME OF THE EVENT. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS NOT REVIEWED. THE LABORATORY'S EXPOSURE CONTROL/RISK MANAGEMENT PLANS ARE IN PLACE. THERE WAS NO BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. THE OPERATOR DID NOT SEEK MEDICAL ATTENTION. THE CUSTOMER INDICATED THAT NO ERRONEOUS RESULTS HAD BEEN REPORTED OUT OF THE LABORATORY WHEN THE LEAK WAS NOTICED. THE CUSTOMER STATED THAT ALL SAMPLES RAN ON THE DAY OF THE EVENT WILL BE RE-RUN ON AN ALTERNATE INSTRUMENT TO CONFIRM THAT NO ERRONEOUS RESULTS WERE GENERATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 340735 | COULTER® LH 750 HEMATOLOGY ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | LH 750 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |