FDA Adverse Event Malfunction Summary report: N

MONOPOLAR CAUTERY INSTRUMENT

MDR report key: 3241644 · Received July 22, 2013

Report

Report Number
2955842-2013-02685
Event Type
Malfunction
Date Received
July 22, 2013
Date of Event
June 25, 2013
Report Date
July 25, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT HAS NOT BEEN RETURNED FOR EVALUATION, THEREFORE, THE ROOT CAUSE OF THE CUSTOMER REPORTED FAILURE MODE CANNOT BE DETERMINED. A FOLLOW-UP MDR WILL BE SUBMITTED IF THE INSTRUMENT IS RETURNED (POST ENGINEERING EVALUATION) OR IF ADDITIONAL INFORMATION IS RECEIVED. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS CONFIRMED THE REPORTED COMPLAINT. THE INSTRUMENT'S ELECTRICAL CONNECTOR WAS FOUND WITH BENT PRONGS AT THE TIP OF THE SNAKE WRIST WHERE THE CAUTERY SPATULA TIP WAS INSTALLED. REPEATED ATTEMPTS TO INSTALL THE SPATULA TIP ON INSTRUMENT FAILED. FAILURE ANALYSIS CONCLUDED THE DAMAGE WAS LIKELY DUE TO MISHANDLING DURING SPATULA INSTALLATION. THE ENDOWRIST® INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SI SURGICAL PROCEDURE THE HOOK WOULD NOT SCREW ON A MONOPOLAR CAUTERY INSTRUMENT. THE INSTRUMENT WAS NOT USED IN THE CASE. THERE WERE NO MISSING PIECES OR FALLEN PIECES REPORTED. NO PATIENT HARM, ADVERSE OUTCOME, OR INJURY WAS REPORTED.