FDA Adverse Event Malfunction Summary report: N

LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR

MDR report key: 3241627 · Received July 22, 2013

Report

Report Number
3015876-2013-00643
Event Type
Malfunction
Date Received
July 22, 2013
Date of Event
July 1, 2013
Report Date
July 1, 2013
Manufacturer
PHYSIO-CONTROL, INC
Product Code
LDD
PMA / PMN Number
K063119
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EXAMINED THE CUSTOMER'S DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO DETERMINED THAT THE CAUSE OF THE REPORTED FAILURE WAS THAT THE POWER PCB TO THERAPY PCB RIBBON CABLE HAD BECOME DISCONNECTED FROM THE POWER SUPPLY. PHYSIO THEN RECONNECTED THE RIBBON CABLE AND AFTER OBSERVING PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING THE UNIT WAS RETURNED TO THE CUSTOMER FOR USE.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WOULD NO LONGER POWER ON USING AC OR DC POWER. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340378 LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES) LDD PHYSIO-CONTROL, INC 20