FDA Adverse Event Malfunction Summary report: N

FOXCROSS PTA CATHETER

MDR report key: 3241608 · Received July 22, 2013

Report

Report Number
2024168-2013-04602
Event Type
Malfunction
Date Received
July 22, 2013
Date of Event
July 2, 2013
Report Date
July 3, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
LIT
PMA / PMN Number
K102705
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE REPORTED CRACK IN THE HUB, INFLATION ISSUES, AND LEAK WERE CONFIRMED. BASED ON A VISUAL AND FUNCTIONAL INSPECTION OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT REVEALED NO OTHER INCIDENTS. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE HUB OF THE FOXCROSS CRACKED; FLUID LEAKED AND THE BALLOON COULD NOT INFLATE. THE TARGET LESION WAS IN THE LEG. REPORTEDLY, THERE WAS NO SIGNIFICANT IMPACT TO THE PATIENT OR A CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340805 FOXCROSS PTA CATHETER PERIPHERAL DILATATION CATHETER LIT AV-TEMECULA-CT 809251

Patients

Seq Age Sex Outcome Treatment
1