FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3241581 · Received July 22, 2013

Report

Report Number
1416980-2013-19363
Event Type
Malfunction
Date Received
July 22, 2013
Date of Event
June 6, 2013
Report Date
June 27, 2013
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS ONE OR MORE CYCLE ADVANCED TO THE NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING THE EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED, WHICH OCCURRED DURING THE THERAPY INITIATED ON (B)(6) 2013 AT 22:58:59. DURING NIGHT DRAIN CYCLE TWO, THE PATIENT'S ULTRAFILTRATION READING WAS 1292ML, INDICATING THE HOME PATIENT (HP) DRAINED 1292ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340429 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO