FDA Adverse Event Malfunction Summary report: N

GYNECARE MORCELLEX TISSUE MORCELLATOR

MDR report key: 3241530 · Received July 22, 2013

Report

Report Number
2210968-2013-14025
Event Type
Malfunction
Date Received
July 22, 2013
Date of Event
June 26, 2013
Report Date
July 9, 2013
Manufacturer
ETHICON INC.
Product Code
HET
PMA / PMN Number
K100280
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM.

Additional Manufacturer Narrative · 1

(B)(4): THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY AND FUNCTIONALLY EVALUATED. UPON EVALUATION, CONTINUOUS SPINNING OF BLADE WAS NOT OBSERVED. THE BLADE STOPPED SPINNING WHENEVER THE TRIGGER SWITCH RELEASED. THE DEVICE OPERATED AS INTENDED DURING EVALUATION.IN ADDITION, A REVIEW OF THE BATCH MANUFACTURING RECORDS WAS CONDUCTED AND THE BATCH MET ALL FINISHED GOODS RELEASE CRITERIA.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC SUPRACERVICAL HYSTERECTOMY ON (B)(6) 2013. DURING THE PROCEDURE, BEFORE INSERTION INTO PATIENT, THE DEVICE WAS PLUGGED IN AND IT WOULD NOT TURN OFF. THERE WAS A CONTINUOUS SPINNING OF THE BLADE. THE DEVICE WAS UNPLUGGED AND PLUGGED BACK IN AT WHICH POINT THE DEVICE WOULD NOT TURN ON. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340424 GYNECARE MORCELLEX TISSUE MORCELLATOR LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES) HET ETHICON INC. UNK MT218127

Patients

Seq Age Sex Outcome Treatment
1