COULTER® LH 780
Report
- Report Number
- 1061932-2013-01528
- Event Type
- Malfunction
- Date Received
- July 22, 2013
- Date of Event
- June 26, 2013
- Report Date
- June 26, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K061616
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
BECKMAN COULTER IS PROVIDING A CORRECTION FOR MDR #: 1061932-2013-01528. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE. THE FSE INSPECTED THE INSTRUMENT AND OBSERVED AIR BUBBLES IN THE RED BLOOD CELL (RBC) DILUENT DISPENSER WHICH HAD CAUSED THE HIGH RBC RESULTS REPORTED. THE FSE REPLACED THE DISPENSER, RESOLVING THE HEMATOCRIT AND HEMOGLOBIN (H&H) ISSUES. THE FSE ALSO OBSERVED DEBRIS ON THE RETIC SHEAR VALVE AND PROCEEDED TO CLEAN THE SHEAR VALVES AND REPLACE THE PILOT ACTUATOR AT VALVE VL89 TO RESOLVE THE ISSUE. THE RBC DISPENSER AND THE SHEAR VALVE ISSUES WERE UNRELATED. CLEANING OF THE SHEAR VALVES, PER IFU, IS NOT A ROUTINE PROCEDURE PERFORMED BY CUSTOMER WITHOUT INSTRUCTIONS FROM A BECKMAN COULTER REPRESENTATIVE. FAILURE MODE WAS RELATED TO BUBBLES IN THE RBC DILUENT DISPENSER (PM2). ANOTHER FAILURE MODE WAS ATTRIBUTED TO THE PILOT ACTUATOR AT VL89.
ALL QUALITY CONTROL (QC) RESULTS WERE WITHIN SPECIFICATION. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE. THE FSE INSPECTED THE INSTRUMENT AND OBSERVED AIR BUBBLES IN THE RBC DILUENT DISPENSER WHICH HAD CAUSED THE HIGH RBC RESULTS REPORTED. THE FSE INDICATED THAT HE REPLACED A DEFECTIVE RED BLOOD CELL (RBC) DILUENT DISPENSER TO ADDRESS ISSUES OF HIGH RBC RESULTS. THE FSE ALSO OBSERVED DEBRIS ON THE RETIC SHEAR VALVE AND PROCEEDED TO CLEAN THE SHEAR VALVES AND REPLACE PILOT ACTUATOR TO RESOLVE THE ISSUE. THE RBC DISPENSER AND THE SHEAR VALVE ISSUES WERE UNRELATED. THE PILOT ACTUATOR WAS NOT IDENTIFIED AS A MALFUNCTION, IT WAS REPLACED AS AN INCIDENTAL MAINTENANCE. FAILURE MODE WAS ATTRIBUTED TO THE BUBBLES IN THE RBC DILUENT DISPENSER. ANOTHER FAILURE MODE WAS RELATED TO MATERIAL BUILDUP ON THE RETIC SHEAR VALVE.
BECKMAN COULTER IS PROVIDING A CORRECTION FOR MDR #: 1061932-2013-01528. THIS FOLLOW-UP REPORT IS TO CORRECT THE ORIGINAL SUBMISSION: THE FILE ATTACHMENT WAS ATTACHED IN ERROR. THE REPORT SHOULD HAVE NO ATTACHMENT.
THE CUSTOMER REPORTED TO BECKMAN COULTER (BEC) THAT THEY OBTAINED HIGHER RED BLOOD COUNT (RBC) RESULTS FROM THE COULTER LH 780 ANALYZER COMPARED WITH RESULTS RECOVERED ON ANOTHER INSTRUMENT. THE CUSTOMER ALSO REPORTED HEMOGLOBIN AND HEMATOCRIT (H AND H ) CHECK FAILURES. PATIENT DATA WAS NOT SUPPLIED. THE FIELD SERVICE ENGINEER (FSE) PROVIDED AN EXAMPLE OF ONE PATIENT RESULTS WHERE THE INSTRUMENT REPORTEDLY RECOVERED A HIGHER RBC RESULTS THAN A RERUN OF THE SAME SAMPLE ON ANOTHER INSTRUMENT. THE ORIGINAL RESULT WAS 4.50 X 10^6 CELLS/¿L, BUT WHEN REPEATED ON ANOTHER INSTRUMENT A RESULT OF 3.75 X 10^6 CELLS/¿L WAS OBTAINED. THE RERUN WAS CONSIDERED CORRECT. THE FSE STATED THAT THERE WERE NO PRINTOUTS AVAILABLE FOR REVIEW AND AS SUCH INSTRUMENT FLAGS AND MESSAGES CANNOT BE REVIEWED. NO ERRONEOUS RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. NO DEATH OR INJURY IS ATTRIBUTED TO THIS EVENT AND THERE WAS NO CHANGE TO PATIENT TREATMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 340968 | COULTER® LH 780 | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | LH780 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |