FDA Adverse Event Malfunction Summary report: N

GYNECARE MORCELLEX TISSUE MORCELLATOR

MDR report key: 3241491 · Received July 22, 2013

Report

Report Number
2210968-2013-14021
Event Type
Malfunction
Date Received
July 22, 2013
Date of Event
July 8, 2013
Report Date
July 9, 2013
Manufacturer
ETHICON INC.
Product Code
HET
PMA / PMN Number
K100280
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). BLADE SEIZED/STOPPED. CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM.

Additional Manufacturer Narrative · 1

(B)(4): THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY AND FUNCTIONALLY EVALUATED. UPON EVALUATION, THE BLADE WAS NOT SEIZED AND THE DEVICE OPERATED AS INTENDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL LAPAROSCOPIC HYSTERECTOMY ON (B)(6) 2013. DURING THE PROCEDURE, THE DEVICE STOPPED WORKING. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340261 GYNECARE MORCELLEX TISSUE MORCELLATOR LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES) HET ETHICON INC. MT217920

Patients

Seq Age Sex Outcome Treatment
1