FDA Adverse Event
Malfunction
Summary report: N
GYNECARE MORCELLEX TISSUE MORCELLATOR
MDR report key: 3241490
·
Received July 22, 2013
Report
- Report Number
- 2210968-2013-14019
- Event Type
- Malfunction
- Date Received
- July 22, 2013
- Date of Event
- July 8, 2013
- Report Date
- July 9, 2013
- Manufacturer
- ETHICON INC.
- Product Code
- HET
- PMA / PMN Number
- K100280
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4): THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY AND FUNCTIONALLY EVALUATED. UPON EVALUATION, THE BLADE WAS NOT SEIZED AND THE DEVICE OPERATED AS INTENDED.
Additional Manufacturer Narrative · 1
(B)(4). BLADE SEIZED/STOPPED. CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM.
Description of Event or Problem · 1
IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL LAPAROSCOPIC HYSTERECTOMY ON (B)(6) 2013. DURING THE PROCEDURE, THE DEVICE STOPPED WORKING. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 341393 | GYNECARE MORCELLEX TISSUE MORCELLATOR | LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES) | HET | ETHICON INC. | UNK | MT217952 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |