FDA Adverse Event Injury Summary report: N

ENERGEN

MDR report key: 3241431 · Received July 22, 2013

Report

Report Number
2124215-2013-11701
Event Type
Injury
Date Received
July 22, 2013
Date of Event
June 28, 2013
Report Date
July 25, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Additional Manufacturer Narrative · 1

DETAILED ANALYSIS IS BEING PERFORMED ON THIS DEVICE. UPON COMPLETION OF ANALYSIS, THIS REPORT WILL BE UPDATED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE WAS EXHIBITING HIGH OUT-OF-RANGE (OOR) IMPEDANCES ON THE LEFT VENTRICULAR (LV) LEAD. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) WAS CONSULTED AND SUGGESTED TRYING THE RIGHT ATRIAL (RA) LEAD IN THE DEVICE, WHICH ALSO GAVE AN OOR IMPEDANCE. IT WAS BELIEVED THERE MIGHT BE SOME CONNECTION ISSUE, SO THE DEVICE WAS EXPLANTED AND SUCCESSFULLY REPLACED. TO DATE, NO ADDITIONAL ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED.

Description of Event or Problem · 1

--

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
341330 ENERGEN IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND N140

Patients

Seq Age Sex Outcome Treatment
1 76 YR Hospitalization| L| R MISMATCH| N160| 0295| 5076| 4554| N140| 4591| 4542